| Classification | Class I Electro-medical Apparatus with type-B part appllied |
| Head | Self-rectifying single-phase enbloc assembly. Tube: KL2 -0.8-70G Toshiba DG-073B-AC |
| Type | 70 kV |
| Rated electrical power | 0,430 kW |
| Supply voltage | 230/240 v (50/60 Hz) single – phase |
| Absorption | 6 A |
| Focal spot | 0,8 mm (IEC 336) |
| Anode current | 8 mA |
| Total filtration | 2 mm Al eq. |
| Fokus-skin distance | 20 cm |
| Ø of long cone at the end of spacer | <6 cm |
| Dispersed radiation | ≤ 0,25 mGy/h a 1meter from focus 1 |
| Selectable anodic current | 4-7 mA |
| Selectable voltage to x-ray tube | 60-70 kVp |
| Intermittent operation | 1 s exposure: 30 s reset max |
| Control logic | WIRELESS deadman |
| Maximum apparent resistance of mains power supply | 0,46 Ω |
| Fuse | Very Fast Acting F 6.3 A H |
| Alluminium filter to be interposed between x-ray window and collimator cone | AL, 1 mm (AIP99,9 UNI3567) |
| Deadman logic control | wireless |